Class II (Laminar Flow) Biosafety Cabinetry
| Cabinet Style A = Bench Unit With Base or Adjustable Legs Provided B = Bench Unit Without Base or Adjustable Legs Provided C = Console |
Window Type H = Hinged S = Sliding F = Fixed |
Bench Height NA = Not Applicable NP = Not Provided |
CBV = Concurrent Balance Value
Due to a change in nomenclature in NSF/ANSI 49 - 2002 “Class II (Laminar Flow) Biosafety Cabinetry,” Class II, Type A cabinets are now referred to as Class II, Type A1 and Class II, Types B3 and A/B3 cabinets as Class II, Type A2. Class II, Types B1 and B2 cabinets remain unchanged.
| Allentown, Inc. | ![]() |
| Route 526 | |
| P.O. Box 698 | |
| Allentown, NJ 08501-0698 | |
| United States | |
| 609-259-7951 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| 34850 Safety Cabinet[1] [2] | A2 | A | 100 - 110 | 50 - 60 | N/A | 4 | 10S | 36 | Casters I.V. Pole U.V. Light |
| 34860 Safety Cabinet[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 6 | 10S | 36 | Casters I.V. Pole U.V. Light |
[1] The first four digits of the serial number refer to date of manufacture.
[2] Beginning with serial number 070570504. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer's operator manual
for appropriate correction factor) without adjusting cabinet airflow balance. Downflow
nominal set-point of 55 fpm was established with I.V. Pole and U.V. Light removed. This
cabinet model was Certified to NSF/ANSI 49-2004.
[3] Beginning with serial number 070570502. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer's operator manual
for appropriate correction factor) without adjusting cabinet airflow balance. Downflow
nominal set-point of 55 fpm was established with I.V. Pole and U.V. Light removed. This
cabinet model was Certified to NSF/ANSI 49-2004.
|
|||||||||
| Baker Company, Inc. (The) | |
| Sanford Airport | |
| P.O. Drawer E | |
| Sanford, ME 04073 | |
| United States | |
| 800-992-2537 | |
| 207-324-8773 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| BCG401[1] | B2 | C | 100 - 110 | 35 - 45 | 664 @ 1.6 | 4 | 8S | 35 | I.V. Pole U.V. Light Casters |
| BCG601[1] | B2 | C | 100 - 110 | 35 - 45 | 993 @ 1.6 | 6 | 8S | 35 | I.V. Pole U.V. Light |
| NCB405[3] | B1 | C | 100 - 110 | 65 - 75 50 - 60 |
276 @ 0.7 | 4 | 8S | NA | I.V. Pole U.V. Light |
| NCB405-INT[3] [4] | B1 | C | 100 - 110 | 65 - 75 50 - 60 |
276 @ 0.7 | 4 | 8S | NA | I.V. Pole U.V. Light |
| NCB605[3] | B1 | C | 100 - 110 | 65 - 75 50 - 60 |
443 @ 0.8 | 6 | 8S | NA | I.V. Pole U.V. Light |
| NCB605-INT[3] [4] | B1 | C | 100 - 110 | 65 - 75 50 - 60 |
443 @ 0.8 | 6 | 8S | NA | I.V. Pole U.V. Light |
| SG 403A HE[2] [5] | A2 | A | 100 - 110 | 45 - 55 42 - 52 |
N/A | 4 | 8S | 37 | I.V. Pole U.V. Light Casters |
| SG 403A HE[2] [6] | A2 | A | 100 - 110 | 47 - 57 40 - 50 |
N/A | 4 | 10S | 37 | I.V. Pole U.V. Light Casters |
| SG 503A HE[2] [7] | A2 | A | 100 - 110 | 53 - 63 46 - 56 |
N/A | 5 | 12S | 37 | Casters I.V. Pole U.V. Light |
| SG 503A HE[2] [8] | A2 | A | 100 - 110 | 50 - 60 48 - 58 |
N/A | 5 | 8S | 37 | Casters I.V. Pole U.V. Light |
| SG 603A HE[2] [9] | A2 | A | 100 - 110 | 47 - 57 40 - 50 |
N/A | 6 | 10S | 37 | I.V. Pole U.V. Light |
| SG 603A HE[2] [9] | A2 | A | 100 - 110 | 47 - 57 40 - 50 |
N/A | 6 | 8S | 37 | I.V. Pole U.V. Light |
[1] Beginning with serial number 103539 for the 4-foot model and serial number 103706 for
the 6-foot model. Inflow nominal setpoint of 105 fpm was established with a direct
airflow reading instrument. This nominal set point was confirmed using the
manufacturer's recommended alternate calculated method with thermal anemometer in a
constricted (3 inch high) access opening (consult manufacturer's operator's manual for
appropriate correction factor), without adjusting cabinet airflow balance. Downflow
nominal set point was established with I.V. Pole and U.V. Light removed. This cabinet
model was Certified to NSF/ANSI 49-2010a. This cabinet qualifies only with the armrest
installed. This cabinet is approved with additional petcocks on the sidewalls and a low
static pressure exhaust plenum. Cabinet models BCG401 INT and BCG601 INT are Certified
for alternate motor/blower combinations and input power supply.
[2] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate calculated method with thermal anemometer in a constricted (3 inch high)
access opening (consult manufacturer's operator's manual for appropriate correction
factor, if applicable).
[3] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate calculated method with thermal anemometer in a constricted (3 inch high)
access opening (consult manufacturer for appropriate correction factor, if applicable),
without adjusting cabinet airflow balance. Zoned downflow nominal set points of 70 fpm
(zone 1 - comprised of the rear and middle rows) and 55 fpm (zone 2 - the row closest
the window opening) were established with I.V. Pole and U.V. Light removed. This
cabinet model was Certified to NSF/ANSI 49-2011. NCBX05 model approved with prefilters
located under the work surface placed over the primary supply filters.
[4] Approved for alternate power modes of 220V/50Hz and 220V/60Hz.
[5] Beginning with serial number 93621. This model qualifies for both uniform and
non-uniform (zoned) downflow. The acceptable uniform downflow range is 45-55 fpm. The
acceptable non-uniform downflow ranges listed above begin with zone 1 (45-55 fpm) at the
rear of the work area (compromised of lateral rear and center rows) and end with zone 2
(42-52 fpm) (comprised of front lateral row) behind the front window enclosing the work
area. Downflow nominal set points were established with the I.V. Pole and U.V. Light
removed. This cabinet model was Certified to NSF/ANSI 49-2007. This cabinet qualifies
only with the armrest installed. Cabinet model number suffixes SP and HE INT are
Certified for alternate motor/blower combinations and input power supply.
[6] Beginning with serial number 93621. This model qualifies for both uniform and
non-uniform (zoned) downflow. The acceptable uniform downflow range is 45-55 fpm. The
acceptable non-uniform downflow ranges listed above begin with zone 1 (47-57 fpm) at the
rear of the work area (compromised of lateral rear and center rows) and end with zone 2
(40-50 fpm) (comprised of front lateral row) behind the front window enclosing the work
area. Downflow nominal set points were established with the I.V. Pole and U.V. Light
removed. This cabinet model was Certified to NSF/ANSI 49-2007. This cabinet qualifies
only with the armrest installed. Cabinet model number suffixes SP and HE INT are
Certified for alternate motor/blower combinations and input power supply.
[7] Beginning with serial number 90883. This model qualifies for both uniform and
non-uniform (zoned) downflow. The acceptable uniform downflow range is 50-60 fpm. The
acceptable non-uniform downflow ranges listed above begins with zone 1 (53-63 fpm) at
the rear of the work area (comprised of lateral rear and center rows) and ending with
zone 2 (46-56 fpm) (comprised of front lateral row) behind the front window enclosing
the work area. Downflow nominal set points were established with the I.V. Pole and U.V.
Light removed. This cabinet model was certified to NSF/ANSI 49-2007. This cabinet
qualifies only with the armrest installed. This model was certified with optional
workarea prefilters installed. ATS autolift with casters is an additional acceptable
option. Cabinet model number suffixes SP and HE INT are certified for alternate motor /
blower combinations and input power supply.
[8] Beginning with serial number 90883. This model qualifies for both uniform and
non-uniform (zoned) downflow. The acceptable uniform downflow range is 50-60 fpm. The
acceptable non-uniform downflow ranges listed above begins with zone 1 (50-60 fpm) at
the rear of the work area (comprised of lateral rear and center rows) and ending with
zone 2 (48-58 fpm) (comprised of front lateral row) behind the front window enclosing
the work area. Downflow nominal set points were established with the I.V. Pole and U.V.
Light removed. This cabinet model was certified to NSF/ANSI 49-2007. This cabinet
qualifies only with the armrest installed. This model was certified with optional
workarea prefilters installed. ATS autolift with casters is an additional acceptable
option. Cabinet model number suffixes SP and HE INT are certified for alternate motor /
blower combinations and input power supply.
[9] Beginning with serial number 93808. This model qualifies for both uniform and
non-uniform (zoned) downflow. The acceptable uniform downflow range is 45-55 fpm. The
acceptable non-uniform downflow ranges listed above begin with zone 1 (47-57 fpm) at the
rear of the work area (comprised of lateral rear and center rows) and end with zone 2
(40-50 fpm) (comprised of front lateral row) behind the front window enclosing the work
area. Downflow nominal set points were established with the I.V. Pole and U.V. Light
removed. This cabinet model was Certified to NSF/ANSI 49-2007. This cabinet qualifies
only with the armrest installed. Cabinet model number suffixes SP and HE INT are
Certified for alternate motor/blower combinations and input power supply.
|
|||||||||
| BIOBASE Biodustry (Shandong) Co., Ltd. | ![]() |
| No. 2717 First Industry Road | |
| Mingshui Economic Development Zone | |
| Zhangqiu, Jinan, Shandong Province | |
| China | |
| 86 531 8121 9800 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| 11228 bbc 86[1] | A2 | A | 100 - 110 | 60 - 70 | N/A | 4 | 8S | 25 | I.V. Pole U.V. Light |
| 11235 bbc 86[2] [3] | A2 | A | 100 - 110 | 60 - 70 | N/A | 3 | 8S | 33 | I.V. Pole |
| 11236 bbc 86[2] [3] | A2 | A | 100 - 110 | 60 - 70 | N/A | 6 | 8S | 33 | I.V. Pole |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set-point of 65 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2009.
[2] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate calculated method with thermal anemometer in a constricted (80 mm high) access
opening (consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set point of 65 fpm was established
with I.V. Pole removed. This cabinet model was Certified to NSF/ANSI 49-2011.
[3] Approved for alternate power modes of 220V/50Hz and 220V/60Hz.
|
|||||||||
| CHC Lab | ![]() |
| 520-1 Youngsan-dong | |
| Yusung-gu | |
| Daejeon 305-500 | |
| Republic of Korea | |
| 82 42 933 0036 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| CHC-777A2-03[1] [3] | A2 | A | 100 - 110 | 65 - 75 | N/A | 3 | 8S | 33 | Casters I.V. Pole U.V. Light |
| CHC-777A2-04[1] [3] | A2 | A | 100 - 110 | 65 - 75 | N/A | 4 | 8S | 33 | Casters I.V. Pole U.V. Light |
| CHC-777A2-06[1] [4] | A2 | A | 100 - 110 | 65 - 75 | N/A | 6 | 8S | 33 | Casters I.V. Pole U.V. Light |
| CHC-888B2-04[2] [4] | B2 | A | 100 - 110 | 60 - 70 | 818 @ 2.4 | 4 | 8S | 33 | I.V. Pole U.V. Light |
| CHC-888B2-06[4] [5] | B2 | A | 100 - 110 | 55 - 65 | 1231@3.16 | 6 | 8S | 32 | I.V. Pole U.V. Light |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's alternate
recomemded method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor), without
adjusting cabinet airflow balance. Downflow nominal set point was established with
I.V.Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2007.
This cabinet qualifies only with the armrest installed.
[2] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's alternate
recomemded method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor), without
adjusting cabinet airflow balance. Downflow nominal set point was established with
I.V.Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2009.
[3] Certified for use with power supplies of 115V/60 Hz, 220V/50 Hz, and 220V/60 Hz.
[4] Certified for use with power supplies of 115V/60 Hz and 220V/60 Hz.
[5] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's alternate
recommended method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor), without
adjusting cabinet airflow balance. Downflow nominal set point was established with
I.V.Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2010.
This cabinet qualifies only with the armrest installed.
|
|||||||||
| Chung Fu Technical Development Co., Ltd. | |
| 7F, No. 687, Mintsu East Road | |
| Taipei 105 | |
| Taiwan | |
| 886 2 2600 1761 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| AHA-143-AA-A[1] | A2 | A | 100 - 110 | 68 - 78 | N/A | 4 | 8S | 26 | I.V. Pole U.V. Light |
| AHA-143-AB-A[1] [2] | A2 | A | 100 - 110 | 68 - 78 | N/A | 4 | 8S | 26 | I.V. Pole U.V. Light |
| AHA-143-BA-A[1] | A2 | A | 100 - 110 | 68 - 78 | N/A | 4 | 8S | 26 | I.V. Pole U.V. Light |
| AHA-143-BB-A[1] [2] | A2 | A | 100 - 110 | 68 - 78 | N/A | 4 | 8S | 26 | I.V. Pole U.V. Light |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's alternate
recomended method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor), without
adjusting cabinet airflow balance. Downflow nominal set point was established with
I.V.Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2009.
[2] Certified for use with power supply of 220V/50Hz.
|
|||||||||
| Esco Micro Pte Ltd | ![]() |
| 21 Changi South Street 1 | |
| 486777 | |
| Singapore | |
| 65 6542 0833 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| LA2-4A1-E[1] [2] | A2 | A | 100 - 110 | 65 - 75 | N/A | 4 | 9S | 34 | I.V. Pole U.V. Light |
| LA2-4A2-E[1] | A2 | A | 100 - 110 | 65 - 75 | N/A | 4 | 9S | 34 | I.V. Pole U.V. Light |
| LA2-4A3-E[1] [3] | A2 | A | 100 - 110 | 65 - 75 | N/A | 4 | 9S | 34 | I.V. Pole U.V. Light |
| LA2-5A1-E[1] [2] | A2 | A | 100 - 110 | 65 - 75 | N/A | 5 | 9S | 36 | I.V. Pole U.V. Light |
| LA2-5A2-E[1] | A2 | A | 100 - 110 | 65 - 75 | N/A | 5 | 9S | 36 | I.V. Pole U.V. Light |
| LA2-5A3-E[1] [3] | A2 | A | 100 - 110 | 65 - 75 | N/A | 5 | 9S | 36 | I.V. Pole U.V. Light |
| LA2-6A1-E[1] [2] | A2 | A | 100 - 110 | 60 - 70 | N/A | 6 | 8S | 35 | I.V. Pole U.V. Light |
| LA2-6A2-E[1] | A2 | A | 100 - 110 | 60 - 70 | N/A | 6 | 8S | 35 | I.V. Pole U.V. Light |
| LA2-6A3-E[1] [3] | A2 | A | 100 - 110 | 60 - 70 | N/A | 6 | 8S | 35 | I.V. Pole U.V. Light |
| LB2-4B2[4] | B2 | A | 100 - 110 | 55 - 65 | 803 @ 1.9 | 4 | 8S | 36 | I.V. Pole U.V. Light |
| LB2-6B2[5] | B2 | A | 100 - 110 | 55 - 65 | 1269 @ 2.1 | 6 | 8S | 36 | I.V. Pole U.V. Light |
| LR2-4S1-E[1] [2] | A2 | A | 100 - 110 | 65 - 75 | N/A | 4 | 9S | 34 | I.V. Pole U.V. Light |
| LR2-4S2-E[1] | A2 | A | 100 - 110 | 65 - 75 | N/A | 4 | 9S | 34 | I.V. Pole U.V. Light |
| LR2-4S3-E[1] [3] | A2 | A | 100 - 110 | 65 - 75 | N/A | 4 | 9S | 34 | I.V. Pole U.V. Light |
| LR2-5S1-E[1] [2] | A2 | A | 100 - 110 | 65 - 75 | N/A | 5 | 9S | 36 | I.V. Pole U.V. Light |
| LR2-5S2-E[1] | A2 | A | 100 - 110 | 65 - 75 | N/A | 5 | 9S | 36 | I.V. Pole U.V. Light |
| LR2-5S3-E[1] [3] | A2 | A | 100 - 110 | 65 - 75 | N/A | 5 | 9S | 36 | I.V. Pole U.V. Light |
| LR2-6S1-E[1] [2] | A2 | A | 100 - 110 | 60 - 70 | N/A | 6 | 8S | 35 | I.V. Pole U.V. Light |
| LR2-6S2-E[1] | A2 | A | 100 - 110 | 60 - 70 | N/A | 6 | 8S | 35 | I.V. Pole U.V. Light |
| LR2-6S3-E[1] [3] | A2 | A | 100 - 110 | 60 - 70 | N/A | 6 | 8S | 35 | I.V. Pole U.V. Light |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set point of 70 fpm was established
with I.V. Pole and U.V. Light removed.This cabinet model was Certified to NSF/ANSI
49-2011.
[2] Certified for use with a power supply of 230V/50Hz.
[3] Certified for use with a power supply of 230V/60Hz.
[4] Beginning with serial number 2005-10005. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer for appropriate
correction factor, if applicable), without adjusting cabinet airflow balance. Downflow
nominal set point was established with I.V. Pole and U.V. Light removed. This cabinet
model was Certified to NSF/ANSI 49-2004. The nominal downflow set point was changed from
70 fpm to 60 fpm during the 5-year recertification test for this model in 2011, starting
with serial number 2011-51874.
[5] Beginning with serial number 2005-9857. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer for appropriate
correction factor, if applicable), without adjusting cabinet airflow balance. Downflow
nominal set point was established with I.V. Pole and U.V. Light removed. This cabinet
model was Certified to NSF/ANSI 49-2004. The nominal downflow set point was changed from
70 fpm to 60 fpm during the 5-year recertification test for this model in 2011, starting
with serial number 2011-56614.
|
|||||||||
| Kewaunee Scientific Corporation | |
| 2700 West Front Street | |
| Statesville, NC 28677 | |
| United States | |
| 704-873-7202 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| INT-1100[1] [2] | A2 | A | 100 - 110 | 60 - 70 | N/A | 3 | 10S | 33 | I.V. Pole U.V. Light |
| INT-1400[2] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 4 | 10S | 33 | I.V. Pole U.V. Light |
| INT-1700[2] [4] | A2 | A | 100 - 110 | 55 - 65 | N/A | 5 | 10S | 33 | I.V. Pole U.V. Light |
| INT-2000[2] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 6 | 10S | 33 | I.V. Pole U.V. Light |
| INT-B1400[5] | B2 | A | 100 - 110 | 55 - 65 | 967 @ 1.7 | 4 | 8S | 27 | I.V. Pole U.V. Light |
| INT-B2000[5] | B2 | A | 100 - 110 | 55 - 65 | 1256 @ 1.9 | 6 | 8S | 34 | I.V. Pole U.V. Light |
[1] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable) without
adjusting cabinet airflow balance. Downflow nominal set-point of 65 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2009.
[2] Certified for use with power supplies of 115V/60Hz and 230V/50Hz.
[3] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable) without
adjusting cabinet airflow balance. Downflow nominal set-point of 55 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model qualifies only with the
armrest installed. This cabinet model was Certified to NSF/ANSI 49-2009.
[4] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable) without
adjusting cabinet airflow balance. Downflow nominal set-point of 60 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model qualifies only with the
armrest installed. This cabinet model was Certified to NSF/ANSI 49-2009.
[5] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable) without
adjusting cabinet airflow balance. Downflow nominal set-point of 60 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model qualifies only with the
armrest installed. This cabinet model was Certified to NSF/ANSI 49-2010a.
|
|||||||||
| Labconco Corporation | ![]() |
| 8845 Prospect | |
| Kansas City, MO 64132 | |
| United States | |
| 816-333-8811 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| 34300 Logic Series[1] [2] | A2 | A | 100 - 110 | 50 - 60 | N/A | 3 | 10S | 36 | I.V. Pole U.V. Light Casters |
| 34308 Logic Series[1] [3] | A2 | A | 100 - 110 | 50 - 60 | N/A | 3 | 8S | 36 | I.V. Pole U.V. Light Casters |
| 34400 Logic Series[2] [4] | A2 | A | 100 - 110 | 50 - 60 | N/A | 4 | 10S | 36 | Casters I.V. Pole U.V. Light |
| 34408 Logic Series[3] [4] | A2 | A | 100 - 110 | 50 - 60 | N/A | 4 | 8S | 36 | Casters I.V. Pole U.V. Light |
| 34410 Logic Series[3] [4] | B2 | A | 100 - 110 | 50 - 60 | 810 @ 1.9 | 4 | 8S | 36 | I.V. Pole U.V. Light |
| 34500 Logic Series[2] [5] | A2 | A | 100 - 110 | 50 - 60 | N/A | 5 | 10S | 36 | Casters I.V. Pole U.V. Light |
| 34508 Logic Series[3] [5] | A2 | A | 100 - 110 | 50 - 60 | N/A | 5 | 8S | 36 | Casters I.V. Pole U.V. Light |
| 34600 Logic Series[2] [6] | A2 | A | 100 - 110 | 50 - 60 | N/A | 6 | 10S | 36 | Casters I.V. Pole U.V. Light |
| 34608 Logic Series[3] [6] | A2 | A | 100 - 110 | 50 - 60 | N/A | 6 | 8S | 36 | Casters I.V. Pole U.V. Light |
| 34610 Logic Series[3] [7] | B2 | A | 100 - 110 | 50 - 60 | 1218 @ 2.2 | 6 | 8S | 36 | Bag-In/ Bag-Out I.V. Pole U.V. Light |
| 34800 PuriCare Series[1] | A2 | A | 100 - 110 | 50 - 60 | N/A | 4 | 10S | 36 | I.V. Pole U.V. Light |
| 34810 PuriCare Series[4] | A2 | A | 100 - 110 | 50 - 60 | N/A | 6 | 10S | 36 | I.V. Pole U.V. Light |
[1] Beginning with serial number 070570504. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer's operator manual
for appropriate correction factor) without adjusting cabinet airflow balance. Downflow
nominal set-point of 55 fpm was established with I.V. Pole and U.V. Light removed. This
cabinet model was Certified to NSF/ANSI 49-2004.
[2] Certified for use with power supplies of 230V/50 Hz and 230V/60 Hz.
[3] Certified for use with power supplies of 100V/50 Hz, 100V/60 Hz, 230V/50 Hz, and 230V/60
Hz.
[4] Beginning with serial number 070570502. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer's operator manual
for appropriate correction factor) without adjusting cabinet airflow balance. Downflow
nominal set-point of 55 fpm was established with I.V. Pole and U.V. Light removed. This
cabinet model was Certified to NSF/ANSI 49-2004.
[5] Beginning with serial number 070570511. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (2.25 inch high) access opening (consult manufacturer's operator manual
for appropriate correction factor) without adjusting cabinet airflow balance. Downflow
nominal set-point of 55 fpm was established with I.V. Pole and U.V. Light removed. This
cabinet model was Certified to NSF/ANSI 49-2007.
[6] Beginning with serial number 070570503. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (4.25 inch high) access opening (consult manufacturer's operator manual
for appropriate correction factor) without adjusting cabinet airflow balance. Downflow
nominal set-point of 55 fpm was established with I.V. Pole and U.V. Light removed. This
cabinet model was Certified to NSF/ANSI 49-2004.
[7] Beginning with serial number 070900000. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (4.25 inch high) access opening (consult manufacturer's operator manual
for appropriate correction factor) without adjusting cabinet airflow balance. Downflow
nomial set-point of 55 fpm was established with I.V. Pole and U.V. Light removed. This
cabinet model was certified to NSF/ANSI 49-2007.
|
|||||||||
NOTE: Bench stand models 3730300, 3730310, 3780200 and 3780100 may be used with cabinet models
34300/34308 Logic Series.
Bench stand models 3730400, 3730410, 3780201 and 3780101 may be used with cabinet models
34400/34408 Logic Series and the 34800 PuriCare Series.
Bench stand models 3730500, 3730510, 3780203, 3780106, 3780107, 3780312, and 3780316 may
be used with cabinet models 34500/34508 Logic Series.
Bench stand models 3730600, 3630610, 3780202 and 3780102 may be used with cabinet models
34600/34608 Logic Series and the 34810 PuriCare Series.
Bench stand model 3730400 may also be used with cabinet models 34410 Logic Series.
All bench stand models beginning with 378 may be used with or without the optional wheel
kit (PN# 3784000).
| Microzone Corporation | |
| 86 Harry Douglas Drive | |
| Ottawa, ON K2S 2C7 | |
| Canada | |
| 613-831-8318 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| BK-2-4[2] [3] | A2 | A | 100 - 110 | 61 - 71 | N/A | 4 | 10H | 27 | U.V. Light I.V. Pole |
| BK-2-4B3[2] | A2 | A | 100 - 110 | 61 - 71 | N/A | 4 | 10H | 27 | U.V. Light I.V. Pole |
| BK-2-4B3-BT[2] | A2 | B | 100 - 110 | 61 - 71 | N/A | 4 | 10H | NA | U.V. Light I.V. Pole |
| BK-2-4BT[2] [3] | A2 | B | 100 - 110 | 61 - 71 | N/A | 4 | 10H | NA | U.V. Light I.V. Pole |
| BK-2-6[1] [3] | A2 | A | 100 - 110 | 61 - 71 | N/A | 6 | 10H | 27 | U.V. Light I.V. Pole |
| BK-2-6B3[1] | A2 | A | 100 - 110 | 61 - 71 | N/A | 6 | 10H | 27 | U.V. Light I.V. Pole |
| BK-2-6B3-BT[1] | A2 | B | 100 - 110 | 61 - 71 | N/A | 6 | 10H | NA | U.V. Light I.V. Pole |
| BK-2-6BT[1] [3] | A2 | B | 100 - 110 | 61 - 71 | N/A | 6 | 10H | NA | U.V. Light I.V. Pole |
[1] Beginning with serial number 802-0954 and all subsequent units. Inflow nominal
set-point of 105 fpm was established with a direct airflow reading instrument. A
corresponding nominal set-point of 118 fpm was confirmed using manufacturer's
recommended alternate method with thermal anemometer above exhaust filter without
adjusting cabinet airflow balance. Downflow nominal set-point of 66 fpm was established
with U.V. Light and I.V. Pole installed. Certified for use with power supply of 220 V /
50 Hz.
[2] Beginning with serial number 801-1693 and all subsequent units. Inflow nominal
set-point of 105 fpm was established with a direct airflow reading instrument. A
corresponding nominal set-point of 122 fpm was confirmed using the manufacturer's
recommended alternate method with thermal anemometer above exhaust filter without
adjusting cabinet airflow balance. Downflow nominal set-point of 66 fpm was established
with U.V. Light and I.V. Pole installed. Certified for use with power supply of 220 V /
50 Hz.
[3] This model has an A2 type design clasification.
|
|||||||||
| Nuaire, Inc. | |
| 2100 Fernbrook Lane | |
| Plymouth, MN 55447 | |
| United States | |
| 763-553-1270 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| NU-425-400[1] [2] [3] | A2 | A | 100 - 110 | 55 - 65 | N/A | 4 | 8S | 32 | I.V. Pole U.V. Light |
| NU-425-400[1] [2] [3] | A2 | A | 100 - 110 | 55 - 65 | N/A | 4 | 10S | 32 | I.V. Pole U.V. Light |
| NU-425-500[1] [3] [4] | A2 | A | 100 - 110 | 55 - 65 | N/A | 5 | 10S | 32 | U.V. Light I.V. Pole |
| NU-425-500[1] [3] [4] | A2 | A | 100 - 110 | 55 - 65 | N/A | 5 | 8S | 32 | U.V. Light I.V. Pole |
| NU-425-600[1] [3] [5] | A2 | A | 100 - 110 | 55 - 65 | N/A | 6 | 8S | 32 | I.V. Pole U.V. Light |
| NU-425-600[1] [3] [5] | A2 | A | 100 - 110 | 55 - 65 | N/A | 6 | 10S | 32 | I.V. Pole U.V. Light |
| NU-427-400[1] [3] [6] | B1 | A | 100 - 110 | 65 - 75 | 282 @ 0.7 | 4 | 8S | 32 | Bag-In/ Bag-Out I.V. Pole U.V. Light |
| NU-427-600[1] [3] [10] | B1 | A | 100 - 110 | 65 - 75 | 474 @ 0.9 | 6 | 8S | 32 | I.V. Pole U.V. Light Bag-In/ Bag-Out |
| NU-430-400[1] [3] [8] [9] | B2 | A | 100 - 110 | 55 - 65 | 785 @ 1.7 | 4 | 8S | 32 | Bag-In/ Bag-Out I.V. Pole U.V. Light |
| NU-430-600[1] [3] [7] [8] | B2 | A | 100 - 110 | 55 - 65 | 1250 @ 1.8 | 6 | 8S | 32 | Bag-In/ Bag-Out I.V. Pole U.V. Light |
| NU-480-400[1] [3] [11] | A2 | A | 100 - 110 | 55 - 65 | N/A | 4 | 8S | 33 | I.V. Pole U.V. Light |
| NU-480-400[1] [3] [11] | A2 | A | 100 - 110 | 55 - 65 | N/A | 4 | 10S | 33 | I.V. Pole U.V. Light |
| NU-480-500[1] [3] [11] | A2 | A | 100 - 110 | 55 - 65 | N/A | 5 | 8S | 33 | I.V. Pole U.V. Light |
| NU-480-500[1] [3] [11] | A2 | A | 100 - 110 | 55 - 65 | N/A | 5 | 10S | 33 | I.V. Pole U.V. Light |
| NU-480-600[1] [3] [11] | A2 | A | 100 - 110 | 55 - 65 | N/A | 6 | 8S | 33 | I.V. Pole U.V. Light |
| NU-480-600[1] [3] [11] | A2 | A | 100 - 110 | 55 - 65 | N/A | 6 | 10S | 33 | I.V. Pole U.V. Light |
| NU-677-400[12] | A2 | A | 100 - 110 | 55 - 65 | N/A | 4 | 12S | 38 | Canopy Connection U.V. Light |
| NU-677-500[13] | A2 | A | 100 - 110 | 55 - 65 | N/A | 5 | 12S | 38 | Canopy Connection U.V. Light |
| NU-677-600[14] | A2 | A | 100 - 110 | 55 - 65 | N/A | 6 | 12S | 38 | Canopy Connection U.V. Light |
[1] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. A corresponding inflow nominal set-point of 105 fpm (corrected to local air
density) was confirmed using manufacturer's recommended alternate method with thermal
anemometer in a constricted (3 inch high) access opening for appropriate correction
factor, if applicable, without adjusting cabinet airflow balance.
[2] Beginning with series 60 and higher. Downflow nominal set-point of 60 fpm was
established 4 inches above the plane of the window with the I.V. Pole and U.V. Light
removed. This cabinet qualifies only with the armrest installed. This cabinet
qualifies with optional DC motor installed. This cabinet model was Certified to
NSF/ANSI 49–2004.
Listed model designation NU 437 with electronic control panel and airflow monitor.
Listed model designation NU 440 with electronic airflow control system.
Listed model may be prefaced with Q, N, L or S (i.e. NU-Q425-400).
[3] Listed model may be suffixed with E. Certified for use with power supplies of 230V/50Hz
and 230V/60Hz.
[4] Beginning with Series 60 and higher. Downflow nominal set-point of 60 fpm was
established 4 inches above the plane of the window with the I.V. Pole and U.V. Light
removed. This cabinet qualifies only with the armrest installed. This cabinet
qualifies with optional DC motor installed. This cabinet model was Certified to
NSF/ANSI 49-2004.
Listed model designation NU 437 with electronic control panel and airflow monitor.
Listed model designation NU 440 with electronic airflow control system.
Listed model may be prefaced with Q, N, L, or S (i.e. NU-Q425-500).
[5] Beginning with series 60 and higher. Downflow nominal set-point of 60 fpm was
established 4 inches above the plane of the window with the I.V. Pole and U.V. Light
removed. This cabinet qualifies only with the armrest installed. This cabinet
qualifies with optional DC motor installed. This cabinet model was Certified to
NSF/ANSI 49–2004.
Listed model designation NU 437 with electronic control panel and airflow monitor.
Listed model designation NU 440 with electronic airflow control system.
Listed model may be prefaced with Q, N, L or S (i.e. NU-Q425-600).
[6] Beginning with Series 60 and higher. Downflow nominal set-point of 70 fpm was
established 4 inches above the plane of the window with the I.V. Pole and U.V. Light
removed. This cabinet model was Certified to NSF/ANSI 49-2011. This cabinet qualifies
with the optional DC motor installed. Listed model may be prefaced with Q, N, L or S
(i.e. NU-Q427-400).
[7] Beginning with series 60 and higher. The downflow nominal set-point of 60 fpm was
established 4 inches above the bottom of the sash with the I.V. Pole and the U.V. Light
removed. This cabinet model qualifies only with the armrest installed. This cabinet
qualifies with the optional DC motor installed. This cabinet model was Certified to
NSF/ANSI 49-2009. Listed model may be prefaced with Q, N, L, or S (i.e. NU-Q430-600).
[8] Listed model designation NU 435 without U.V. Light option.
Listed model may be prefaced with Q, N, L or S (i.e. NU-Q435-400).
[9] Beginning with series 60 and higher. The downflow nominal set point of 60 fpm was
established 4 inches above the bottom of the sash with the I.V. Pole and U.V. Light
removed. This cabinet model qualifies only with the armrest installed. This cabinet
qualifies with the optional DC motor installed. This cabinet model was Certified to
NSF/ANSI 49-2009. Listed model may be prefaced with Q, N, L, or S (i.e. NU-Q430-400).
[10] Beginning with series 40 and higher. Downflow nominal set-point of 60 fpm was
established 4 inches above the plane of the window with the I.V. Pole and U.V. Light
removed. This cabinet qualifies with the optional DC motor installed.
Listed model may be prefaced with Q, N, L or S (i.e. NU-Q427-600). This cabinet model
was Certified to NSF/ANSI 49-2011.
[11] Beginning with series 1 and higher. The downflow nominal set-point of 60 fpm was
established 4 inches above the bottom of the sash with the I.V. Pole and U.V. Light
removed. This cabinet model was Certified to NSF/ANSI 49-2009. Listed model may be
prefaced with Q, N, L, or S (i.e. NU-Q480-400). Listed model designation NU-475 with
analog control panel. Listed model designation NU-477 with electric control panel and
airflow monitor.
[12] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. A corresponding inflow nominal set-point of 105 fpm (corrected to local air
density) was confirmed using manufacturer's recommended alternate method with thermal
anemometer in a constricted (3 inch high) access opening and applying appropriate
correction factor without adjusting cabinet airflow balance. The downflow nominal
set-point of 60 fpm was established 4 inches above the bottom of the sash with the U.V.
Light removed. This cabinet model was Certified to NSF/ANSI 49-2011. NuAire canopy
model NU-911-402 was evaluated as part of the NSF Certification testing procedures.
[13] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. A corresponding inflow nominal set-point of 105 fpm (corrected to local air
density) was confirmed using manufacturer's recommended alternate method with thermal
anemometer in a constricted (3 inch high) access opening and applying appropriate
correction factor without adjusting cabinet airflow balance. The downflow nominal
set-point of 60 fpm was established 4 inches above the bottom of the sash with the U.V.
Light removed. This cabinet model was Certified to NSF/ANSI 49-2011. NuAire canopy
model NU-911-502 was evaluated as a part of the NSF Certification testing procedures.
[14] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. A corresponding inflow nominal set-point of 105 fpm (corrected to local air
density) was confirmed using manufacturer's recommended alternate method with thermal
anemometer in a constricted (3 inch high) access opening and applying appropriate
correction factor without adjusting cabinet airflow balance. The downflow nominal
set-point of 60 fpm was established 4 inches above the bottom of the sash with the U.V.
Light removed. This cabinet model was Certified to NSF/ANSI 49-2011. NuAire canopy
model NU-911-602 was evaluated as part of the NSF Certification testing procedures.
|
|||||||||
| Thermo Fisher Scientific | |
| Robert-Bosch-Str. 1 | |
| D-63505 Langenselbold | |
| Germany | |
| 49 61 813 5277 | |
| Visit this company's website |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| 1305[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1306[1] [2] | A2 | A | 100 - 110 | 58 - 58 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1307[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1308[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1320[5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1321[5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1322[5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1323[5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest U.V. Light I.V. Pole |
| 1325[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1326[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1327[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1328[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1330[2] [5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1331[2] [5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1332[2] [5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1333[2] [5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1335[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1336[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1337[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1338[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1339[5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1340[5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1341[5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1342[5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1345[4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1346[4] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1347[4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1348[4] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1350[5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1351[5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1352[5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1353[5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1354[3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1355[4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1356[4] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1357[4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1358[4] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1359[4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1366[2] [5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1367[2] [5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1368[5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1369[5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest U.V. Light I.V. Pole |
| 1370[5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1371[5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1372[2] [5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest U.V. Light I.V. Pole |
| 1373[2] [5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1375[4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1376[4] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1377[4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1378[4] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1380[2] [5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1381[2] [5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1382[2] [5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1383[2] [5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 5 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1384[3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1385[4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1386[4] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1387[4] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1388[4] [7] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1389[3] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1390[2] [5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1391[2] [5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1392[2] [5] [6] | A2 | B | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1393[2] [5] [6] | A2 | A | 100 - 110 | 58 - 68 | N/A | 3 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1395[7] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1396[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1397[7] [8] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
| 1398[1] [2] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 8S | 36 | Armrest I.V. Pole U.V. Light |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2004.
[2] Certified for use with power supply of 230V/50Hz and 230V/60Hz.
[3] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2004. Certified for use with power supply of 220V/50Hz.
[4] Inflow nominal set-point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2004.Certified for use with power supply of 230V/50Hz and 230V/60Hz.
[5] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set point was confirmed using the manufacturer's alternate
recomended method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer's operator's manual for appropriate correction factor), without
adjusting cabinet airflow balance. Downflow nominal set point was established with
I.V.Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI 49-2009.
[6] May be used with alternate, adjustable height electric stand (maximum bench height of 35
inches).
[7] Certified for use with power supply of 220 V / 50 Hz and 220 V / 60 Hz.
[8] Beginning with serial number 188122-1 and 188108-1 for models 1395 and 1397,
respectively. Inflow nominal set-point of 105 fpm was established with a direct airflow
reading instrument. This nominal set point was confirmed using the manufacturer's
recommended alternate method with thermal anemometer in a constricted (3 inch high)
access opening (consult manufacturer for appropriate correction factor, if applicable),
without adjusting cabinet airflow balance. Downflow nominal set point of 63 fpm was
established with I.V. Pole and U.V. Light removed. This cabinet model was Certified to
NSF/ANSI 49-2009.
|
|||||||||
Facility : Shanghai, China
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| 1354[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1359[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1384[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 4 | 10S | 36 | Armrest I.V. Pole U.V. Light |
| 1389[1] | A2 | A | 100 - 110 | 58 - 68 | N/A | 6 | 10S | 36 | Armrest I.V. Pole U.V. Light |
[1] Inflow nominal set point of 105 fpm was established with a direct airflow reading
instrument. This nominal set-point was confirmed using the manufacturer's recommended
alternate method with thermal anemometer in a constricted (3 inch high) access opening
(consult manufacturer for appropriate correction factor, if applicable), without
adjusting cabinet airflow balance. Downflow nominal set-point of 63 fpm was established
with I.V. Pole and U.V. Light removed. This cabinet model was Certified to NSF/ANSI
49-2004. Certified for use with power supply of 220V/50Hz.
|
|||||||||
Facility : Langenselbold, Germany
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| KS12[1] | A2 | A | 100 - 110 | 70 - 80 | N/A | 4 | 10 10H S |
34 | Full Side Windows U.V. Light I.V. Pole Armrest Casters |
| KS15[4] | A2 | A | 100 - 110 | 65 - 75 | N/A | 5 | 10 10H S |
35 | Full Side Windows I.V. Pole U.V. Light Armrest Casters |
| KS18[2] | A2 | A | 100 - 110 | 65 - 75 | N/A | 6 | 10 10H S |
34 | Full Side Windows I.V. Pole U.V. Light Casters |
| KS9[3] | A2 | A | 100 - 110 | 65 - 75 | N/A | 3 | 10 10H S |
35 | Full Side Windows I.V. Pole U.V. Light Armrest Casters |
[1] Beginning with serial number 40292176. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. A corresponding inflow nominal
set-point of 105 fpm (corrected to local air density) was confirmed using the
manufacturer's recommended alternate method with thermal anemometer in a constricted (3
inch high) access opening (consult manufacturer's operator's manual for appropriate
correction factor, if applicable) without adjusting cabinet airflow balance. This
cabinet model was Certifed with a 4 section work surface. This cabinet model was
Certified to NSF/ANSI 49-2002.
[2] Beginning with serial number 40292176. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. A corresponding inflow nominal
set-point of 105 fpm (corrected to local air density) was confirmed using the
manufacturer's recommended alternate method with thermal anemometer in a constricted (3
inch high) access opening (consult manufacturer's operator's manual for appropriate
correction factor, if applicable) without adjusting cabinet airflow balance. This
cabinet model was Certified with a 6 section work surface. This cabinet model was
Certified to NSF/ANSI 49-2002.
[3] Beginning with serial number 40329642. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer's operator's manual
for appropriate correction factor) without adjusting cabinet airflow balance. This
cabinet model was Certified with a 3 section work surface. This cabinet model was
Certified to NSF/ANSI 49-2002.
[4] Beginning with serial number 40349384. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer's operator's manual
for appropriate correction factor) without adjusting cabinet airflow balance. This
cabinet model was Certified with a 5 section work surface. This cabinet model was
Certified to NSF/ANSI 49-2002.
|
|||||||||
| Thermo Fisher Scientific | |
| 401 Millcreek Road | |
| Marietta, OH 45750 | |
| United States | |
| 740-373-4763 |
| Model Number |
Cabinet Type/Style |
Inflow Velocity (fpm) |
Downflow Velocity (fpm) |
CBV (cfm) at Static Pressure (in w.g.) |
Cabinet Width ft. |
Window Ht/Type in. |
Bench Ht Max in. |
Acceptable Options |
|
| 1310[1] [2] | B2 | A | 100 - 110 | 50 - 60 | 810 @ 1.9 | 4 | 8S | 36 | Bag-In/ Bag-Out I.V. Pole U.V. Light |
| 1360[3] [4] | B2 | A | 100 - 110 | 50 - 60 | 1218 @ 2.2 | 6 | 8S | 36 | Bag-In/ Bag-Out I.V. Pole U.V. Light |
[1] Beginning with serial number 070570502. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (3 inch high) access opening (consult manufacturer's operator manual
for appropriate correction factor) without adjusting cabinet airflow balance. Downflow
nominal set-point of 55 fpm was established with I.V. Pole and U.V. Light removed. This
cabinet model was Certified to NSF/ANSI 49-2004. Model 1311 is equivalent to Model
1310, but comes equipped with U.V. lighting (Listed as an acceptable option).
[2] Certified for use with alternate power supplies. Model 1315 is equivalent to model
1310, but is rated for use with power supplies of 230V and 50/60 Hz and is equipped with
a Chinese-compatible plug. Model 1316 is equivalent, but is rated for use with power
supplies of 230V and 50/60 Hz and is equipped with U.V. lighting and a
Chinese-compatible plug. Model 1317 is equivalent, but is rated for use with power
supplies of 230V and 50/60 Hz and is equipped with a North American 230 Volt-compatible
plug. Model 1318 is equivalent, but is rated for use with power supplies of 230V and
50/60 Hz and is equipped with U.V. lighting and a North American 230 Volt-compatible
plug.
[3] Beginning with serial number 070900000. Inflow nominal set-point of 105 fpm was
established with a direct airflow reading instrument. This nominal set-point was
confirmed using the manufacturer's recommended alternate method with thermal anemometer
in a constricted (4.25 inch height) access opening (consult manufacturer’s operator
manual for appropriate correction factor) without adjusting cabinet airflow balance.
Downflow nominal set-point of 55 fpm was established with I.V. Pole and U.V. Light
removed. This cabinet model was Certified to NSF/ANSI 49-2007. Model 1361 is
equivalent to Model 1360, but is equipped with U.V. lighting (listed as an acceptable
option).
[4] Certified for use with alternate power supplies. Model 1362 is equivalent to model
1360, but is rated for use with power supplies of 230V and 50/60 Hz and is equipped with
a Chinese-compatible plug. Model 1363 is equivalent, but is rated for use with power
supplies of 230V and 50/60 Hz and is equipped with U.V. lighting and a
Chinese-compatible plug. Model 1364 is equivalent, but is rated for use with power
supplies of 230V and 50/60 Hz and is equipped with a North American 230 Volt-compatible
plug. Model 1365 is equivalent, but is rated for use with power supplies of 230 V and
50/60 Hz and is equipped with U.V. lighting and a North American 230 Volt-compatible
plug.
|
|||||||||
Number of matching Manufacturers is 12